Skip to content

Public FDA record

Z-1703-2008: Circular Mapping Catheter-Biosense Webster LASSO 2515 Variable Circular Mapping Catheter- The catheter is

Recall record Z-1703-2008 for Biosense Webster Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1703-2008
Company in FDA record
Biosense Webster Inc
FDA classification date
Aug 25, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This recall was initiated after the firm received several complaints reporting issues with deflection. The catheter may become caught in the mitral valve, requiring surgical procedures, valve may be torn while trying to dislodge, unable to retract the catheter through the sheath and require surgical intervention to remove, pulling on catheter to remove may lead to large atrial septal tear, and sep

Source facts

Action
A letter entitled, "Urgent Safety Field Notice" dated March 25, 2008, and certification form were sent to all customers and Johnson & Johnson affiliates who purchased the product in the past year as of March 25, 2008, due to the expiration date. For customers in the US, two letters and forms were sent out - one addressed to the EP/Cardiology Lab and a second to the Risk Management Office. If you have questions or have not received a letter contact Customer Service at 1-800-729-9010.
Initiated
2008-03-25
Root cause
Device Design
Terminated
2012-05-03
Distribution
Nationwide and Worldwide to: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czechia, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary , Iran, Ireland, Isra¿¿l, Italy, Japan, Lithuania, Luxembourg, Mexico, Middle East, Netherlands, Norway, Panama, Poland, Portugal, Russia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom
Recall number
Z-1703-2008
Recall status
Terminated
CDRH record ID
69666
FDA classified
2008-08-25
Recall event number
47522

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.