The Surgical Protocol has been modified to create separate and distinct surgical protocols, one for the Trident PSL Shell (LSP68) and one for the Trident Hemispherical Shell (LSP69) in order to clarify and highlight the differences in reaming technique.
Source facts
Action
Stryker sent Urgent - Product Correction - Surgical Protocol letters by Federal Express on June 9, 2009 to all Stryker branches/agencies, hospital Risk Management, hospital Chief of Orthopaedics, surgeons and all foreign consignees.
Initiated
2009-06-09
Root cause
Labeling design
Terminated
2012-05-08
Distribution
Worldwide distribution.
Recall number
Z-1703-2009
Recall status
Terminated
CDRH record ID
82672
FDA classified
2009-08-04
Recall event number
52289
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.