Open original FDA record ↗Recall record
- FDA identifier
- Z-1703-2017
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Mar 19, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens Healthcare Diagnostics has confirmed that in isolated cases when ECREA is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of greater than 15 percent in the ECREA result.
Source facts
- Action
- Siemens mailed an Urgent Medical Device Correction Notice letter dated March 3, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Siemens Customer Care Center - Technical Solutions at (800-441-9250) or your local Siemens Technical Support Representative.
- Initiated
- 2017-03-03
- Root cause
- Under Investigation by firm
- Terminated
- 2018-07-03
- Distribution
- Worldwide Distribution - United States Nationwide and Puerto Rico, Canada, Qatar, Republic Korea, Saudi Arabia, Slovakia, Slovenia, Spain, Switzerland, United Kingdom, Australia, Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, and Portugal.
- Recall number
- Z-1703-2017
- Recall status
- Terminated
- CDRH record ID
- 153908
- FDA classified
- 2017-03-19
- Recall event number
- 76690