Z-1703-2018: ATTUNE Revision CRS Femoral LT SZ 5 Cemented, Catalog No. 150440105, Femoral Revision Implant, Knee

Recall record Z-1703-2018 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1703-2018
Company in FDA record
DePuy Orthopaedics, Inc.
FDA classification date
Apr 30, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There may be burrs on the extraction hole threads.

Source facts

Action
Notification letters distributed on 4/17/18 instructed customers to perform the following: Steps to Take Please take the following actions: " Immediately return unused affected units: Note: On March 19, 2018, one unit was removed from consignment. Two units were implanted prior to the medical device recall. No additional units are in the market. " Reconciliation Form: Complete the reconciliation form and return to your sales consultant or fax to 574-371-4939 or email to klong16@its.jnj.com within five (5) days of this notice. " Records: Retain a copy of the completed reconciliation form in your files along with this notice. " Additional Notifications: o Notify surgeons at your facility by providing them with a copy of this notice. Forward this notice to others in your facility that need to be informed. Clinical Implications and Patient Impact DePuy (Ireland) is not recommending prophylactic revision in the absence of symptoms. The company recommends that surgeons discuss potential clinical implications and risks with symptomatic patients that received the affected units.
Initiated
2018-04-16
Root cause
Process control
Terminated
2019-04-02
Distribution
The products were distributed to the following US states: FL, MO, and PA.
Recall number
Z-1703-2018
Recall status
Terminated
CDRH record ID
163577
FDA classified
2018-04-30
Recall event number
79818

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗