Open original FDA record ↗Recall record
- FDA identifier
- Z-1703-2020
- Company in FDA record
- Sysmex America, Inc.
- FDA record creation date
- Apr 14, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.
Source facts
- Action
- Sysmex America, Inc. notified customers on 03/09/2020 via "Product Notification SP-50 and CF-70 Software Mismatch Field Correction" letter. The recall letter identified the affected product and model numbers. The customers were asked to verify the software version. If the versions did not match, a service visit is required for the correction of the software mismatch.
- Initiated
- 2020-03-09
- Root cause
- Software design
- Terminated
- 2021-03-24
- Distribution
- USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..
- Recall number
- Z-1703-2020
- Recall status
- Terminated
- CDRH record ID
- 180368
- FDA record created
- 2020-04-14
- Recall event number
- 85254