Skip to content

Public FDA record

Z-1703-2020: The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stain

Recall record Z-1703-2020 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1703-2020
Company in FDA record
Sysmex America, Inc.
FDA record creation date
Apr 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.

Source facts

Action
Sysmex America, Inc. notified customers on 03/09/2020 via "Product Notification SP-50 and CF-70 Software Mismatch Field Correction" letter. The recall letter identified the affected product and model numbers. The customers were asked to verify the software version. If the versions did not match, a service visit is required for the correction of the software mismatch.
Initiated
2020-03-09
Root cause
Software design
Terminated
2021-03-24
Distribution
USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..
Recall number
Z-1703-2020
Recall status
Terminated
CDRH record ID
180368
FDA record created
2020-04-14
Recall event number
85254

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗