Z-1703-2025: myQA iON; Article Number: MQ10-000;

Recall record Z-1703-2025 for IBA Dosimetry GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1703-2025
Company in FDA record
IBA Dosimetry GmbH
FDA classification date
Apr 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Wrong analysis results can occur if the user creates two structures with the same name, except for one differing by a special character.

Source facts

Action
On April 1, 2025, FIELD SAFETY NOTICE letters were sent to customers. Actions to be taken to prevent the error from occurring: When naming structures, please refrain from using any of the following special characters: % \ / : * ? < > | + Additionally, during the export of files into myQA iON, some TPSs cut off the structure names due to their length. Users are not notified when names are automatically shortened by the system. Please check that the structure name was correctly imported. Distribution of this Field Safety Notice: Please distribute this notice to all those who need to be aware within your area of responsibility where the affected medical device is in use. We would kindly ask you to acknowledge the receipt of this Field Safety Notice. Please sign and send back the attached Confirmation of Receipt within four weeks after the receipt of this Field Safety Notice. Contact information: Please feel free to contact us as follows if you have any further questions: USA, Canada, Latin America Europe, Middle East, Africa Asia Pacific Phone: +1 786 288 0369 Phone: +49 9128 607 38 Phone: +65 3129 2472 service-usa@iba-group.com service-emea@iba-group.com service-apac@iba-group.com
Initiated
2025-04-01
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, FL, GA, KS, MD, OK and the countries of Algeria, Austria, Belgium, Czech Republic, Denmark, France, Germany, Ireland, Italy, Morocco, Norway, Poland, Slovakia, Spain, Israel, Kazakhstan, Kosovo, Libya, South Africa, Turkey, United Kingdom, Switzerland, Argentina, Brazil, Mexico, APAC: Indonesia, Japan, Peoples Republic of China, Philippines, South Korea, Thailand.
Recall number
Z-1703-2025
Recall status
Open, Classified
CDRH record ID
213352
FDA classified
2025-04-30
Recall event number
96647

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗