Open original FDA record ↗Recall record
- FDA identifier
- Z-1704-2010
- Company in FDA record
- Intelligent Hearing Systems Corp.
- FDA classification date
- May 27, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This was a marketing correction mandated by the United States Food and Drug Administration (USFDA). USFDA notified Intelligent Hearing Systems that 501(k) premarket notification application must be submitted and cleared in order to market and sell devices with Vestibular Evoked Myogenic Potential (VEMP) specific functions. The Food and Drug Administration considers that the VEMP testing functions
Source facts
- Action
- Intelligent Hearing Systems sent a letter "Information Regarding the Discontinued Use of the SmartEP Evoked Response System for Vestibular Evoked Myogenic Potentials (VEMP) Testing" dated September 30, 2008 to customers and distributors. Each customer will receive an upgrade and VEMP removal kit which will remove the VEMP feature.
- Initiated
- 2008-10-21
- Root cause
- Device Design
- Terminated
- 2011-09-21
- Distribution
- Nationwide distribution.
- Recall number
- Z-1704-2010
- Recall status
- Terminated
- CDRH record ID
- 87751
- FDA classified
- 2010-05-27
- Recall event number
- 54168