Open original FDA record ↗Recall record
- FDA identifier
- Z-1704-2012
- Company in FDA record
- Invacare Corporation
- FDA classification date
- Jun 5, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Adverse event data prompted testing of device, which revealed that the rod supporting the knee pad was not made to specifications. Upon placing a load on the knee pad, the rod may break or bend causing the walker to collapse. As a result, the user may fall.
Source facts
- Action
- Invacare notified all consignees and distributors with a "Urgent: Recall Information" letter dated May 16, 2012. The product issue was described and recommended actions were provided to customers. Questions were directed to 1-888-257-7912.
- Initiated
- 2012-04-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-01
- Distribution
- Worldwide Distribution--USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of China.
- Recall number
- Z-1704-2012
- Recall status
- Terminated
- CDRH record ID
- 108408
- FDA classified
- 2012-06-05
- Recall event number
- 61550