Open original FDA record ↗Recall record
- FDA identifier
- Z-1705-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jun 3, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
measurements drawn on the 2nd and subsequent images of the series are not visible on printouts when the series is sent to print.
Source facts
- Action
- Siemens sent a Customer Advisory Notice, dated May 15, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to check the software version of their syngo.via system and consider whether the issue described in the advisory notice impact their clincal coutine and whether to arrange the installation of SY027/15/P (VA30A_HF07). Customers were advised to include the advisory notice in their operator's manual. Customers were also advised to notify anyone at their organization who should be aware of this information. For questions regarding this recall call 610-219-6300.
- Initiated
- 2015-05-15
- Root cause
- Software design
- Terminated
- 2017-03-24
- Distribution
- Nationwide Distribution
- Recall number
- Z-1705-2015
- Recall status
- Terminated
- CDRH record ID
- 137299
- FDA classified
- 2015-06-03
- Recall event number
- 71314