Open original FDA record ↗Recall record
- FDA identifier
- Z-1705-2016
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- May 23, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If by mistake the initial pre-treatment imaging was performed BEFORE the HexaPOD was moved to the DRIVE (*START) position, iGUIDE offers the possibility to branch into a specific catch-up workflow, which was implemented to use the initial scan data.
Source facts
- Action
- Elekta sent an Important Field Safety Notice that was distributed to all affected customers on May 12, 2016. The notice informs end users about the problem and the potential clinical impact. It also gives a recommendation how the risk can be mitigated. The notice also advises the customer that a patch will be released that will introduce a patch that avoids an inaccurate position and thus eliminates an unsafe situation. Corrective Action #2: Permanent Solution Software Upgrade A software patch will be released that will correct the behavior of the iGUIDE software. The target release date for this patch is January 2017. Service teams will have 6 months from the date of release to correct all affected devices in the field. For further questions, call (770) 300-9725.
- Initiated
- 2016-05-12
- Root cause
- Software design
- Terminated
- 2021-07-16
- Distribution
- IL, LA, Austria, Australia, Botswana, Germany, Denmark, France, India, Japan
- Recall number
- Z-1705-2016
- Recall status
- Terminated
- CDRH record ID
- 146250
- FDA classified
- 2016-05-23
- Recall event number
- 74087