Skip to content

Public FDA record

Z-1705-2018: Lactosorb RapidFlap, bone plate, Model No. 915-0020 The RapidFlap LS Cranial Fixation System is indicate

Recall record Z-1705-2018 for Zimmer Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1705-2018
Company in FDA record
Zimmer Biomet, Inc.
FDA classification date
May 7, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recalling firm has confirmed that the Outer Plate component exhibits an excessive chamfer on the threading after deburring operations. This excessive chamfer results in non-conforming product where the threads of the outer plate component have limited to no engagement with the post component.

Source facts

Action
Zimmer Biomet sent an Urgent Medical Device Recall letter dated April 5, 2018, to Distributors, Sales Representatives, and Distribution Managers. A separate recall letter was sent to Risk Managers on the same date. (Updated 3/20/19) On February 12, 2019, the firm issued notification of recall expansion. Further affected lots had been identified as part of the recall, and affected customers were notified. (/Update) Consignees were informed of the product issue and risk of non-functioning device and delays in surgery. All parties were instructed to review the notice, locate and quarantine affected product in inventory, return all affected product with a completed return form and mark "RECALL" on the outside of returned cartons, and return the completed customer response form. If you have further questions or concerns regarding this recall, please call (904)741-4400 extension 9133 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of operating hours will receive a voicemail prompt. For further questions, please call (574) 371-3071. (Updated 3/20/19) Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. (/Update)
Initiated
2018-04-05
Root cause
Process control
Terminated
2020-08-12
Distribution
worldwide Distribution - US Distribution to the states of : CA, FL, MO, NC, SC, TX, and WI., and to the countries of : Argentina, China, Colombia, Germany, Japan, Netherlands, South Africa, Thailand, and Vietnam.
Recall number
Z-1705-2018
Recall status
Terminated
CDRH record ID
164133
FDA classified
2018-05-07
Recall event number
79987

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗