Z-1705-2019: IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.4" (100 cm), Rx Only, Steril

Recall record Z-1705-2019 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1705-2019
Company in FDA record
Edwards Lifesciences, LLC
FDA classification date
Jun 11, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the operative site may be obscured, and the device will need to be exchanged or operative strategy would need to change, including placement of an external cross-clamp, conversion to an open procedure, or performing the procedure under fibrillation.

Source facts

Action
On 05/14/19, the firm, Edwards, mailed "Urgent Product Recall" notices, via overnight service to its customers. The letter described the product, problem and actions to be taken. The letter informed customers of the recall and instructed them to do the following: 1) Review this field safety notice to understand the potential hazard. 2) Complete and return the attached Acknowledgement Form within five (5) business days of receiving this notice to Customer Service. 3) Complete and return the Product Reconciliation Form to Customer Service; return affected devices with the Return Goods Authorization (RGA) provided. 4) Distribute this notice within your organization or to any organization where the potentially affected devices have been transferred. If you have further distributed this product, notify your customers to the user level. Report any balloon failures to the recalling firm. Customers with additional questions were encouraged to call Customer Service Monday through Friday at (800) 424-3278, Option 1, from 8:00 AM - 4:00 PM Pacific Time. On 07/09/19, Urgent Product Recall notices were mailed via next-day service to customers informing them of the recall expansion.
Initiated
2019-05-14
Root cause
Device Design
Terminated
2024-04-23
Distribution
Worldwide Distribution: US (nationwide) to states of: UT, GA, NJ, CA, OH, PA, AL, FL, TN, TX, ME, MA, IN, MS, NY, VA, MI, IL, CO, OR, WA, SC, AZ, SD, MO, CT, OK, NV, WV, WI, NC, MN, KY; and countries of: United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Czech Republic, Germany, France, United Kingdom, Italy, Netherlands, Poland, Romania, and Sweden.
Recall number
Z-1705-2019
Recall status
Terminated
CDRH record ID
172949
FDA classified
2019-06-11
Recall event number
82870

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗