When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image.
Source facts
Action
Philips Medicals Systems distributed an "URGENT Field Safety Notice" alerting customers of the defect on November 28, 2012. Customers are requested to manually rotate images by using the rotation tool.Customers are also recommended to check whether the automatic marker is set correctly, especially after a manual image rotation is performed. The marker can be corrected manually with the annotation tool.
Initiated
2012-11-28
Root cause
Software design
Terminated
2018-05-31
Distribution
Nationwide (AR, AZ, CA, DE, FL, GA, IA, IL, IN, KS, MD, MI, MN, NC, NM, NY, OH, OR, PA, SC, TX, VT, WI)
Recall number
Z-1706-2017
Recall status
Terminated
CDRH record ID
117359
FDA record created
2017-04-04
Recall event number
64902