This lot was packaged with the wrong occluding ball guidewire. The PulseSpray Catheter is 90 cm in length and requires a 109 cm occluding ball guidewire, however this lot was packaged with a 60 cm occluding ball guidewire.
Source facts
Action
Angiodynamics issued an "Urgent Medical Device Recall" notification (dated 5/11/10) to user accounts via Certified Mail. Consignees were asked to identify, segregate and return all affected product to the firm in addition to sending in a completed Reply Form via fax.
For further information, contact Angiodynamics Customer Service at 1-800-772-6446.
Initiated
2010-05-17
Root cause
Process control
Terminated
2012-07-25
Distribution
Worldwide Distribution -- United States (AR, NY, PA, MI, MN, LA and WA), Spain, Austria and Denmark.
Recall number
Z-1707-2010
Recall status
Terminated
CDRH record ID
91638
FDA classified
2010-05-27
Recall event number
55715