Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to
Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]
Source facts
Action
Philips issued Field Safety Notice on 3/6/17. The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer .Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice:
When a No Data Tele INOP alert is displayed on a Philips IntelliVue Information Center iX (PIIC iX), check the MX40 as directed in the PIIC iX B.01 IFU, page 87. If the MX40 is in the Coverage Area, remove the battery, and reinsert the battery. If the MX40 is not in the Coverage Area, place the sector in the PIIC iX in standby as directed in the MX40 B.06 IFU, page 53, or return the patient and MX40 to the Coverage Area.
Philips is releasing a software upgrade B.06.18 for affected Philips IntelliVue MX40 Patient Wearable Monitors.
A Philips Healthcare representative will contact customers with affected devices to arrange for the installation of the software upgrade
Initiated
2017-03-06
Root cause
Software design
Terminated
2018-05-25
Distribution
Worldwide Distribution - US Nationwide including the countries of:
Australia
Belgium
Canada
Denmark
Finland
France
Germany
Germany
India
Italy
Netherlands
Norway
Qatar
Singapore
South Africa
Sweden
Switzerland
United Kingdom
Recall number
Z-1707-2017
Recall status
Terminated
CDRH record ID
153762
FDA record created
2017-04-04
Recall event number
76656