Skip to content

Public FDA record

Z-1708-2020: Positron Emission Tomograpy and Computed Tomography System, Model: uMI 550, with software version R001.3.

Recall record Z-1708-2020 for Shanghai United Imaging Healthcare Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1708-2020
Company in FDA record
Shanghai United Imaging Healthcare Co., Ltd.
FDA record creation date
Apr 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential sporadic software bugs in R001.3.0.0.750505 software version may cause an issue with the PET acquisition raw data causing reconstruction problems during PET/CT scanning and could possibly cause rescans of patients with additional dose.

Source facts

Action
On August 26, 2019, the firm emailed consignee an "Customer Notification" recall letter. In addition to informing consignees about the recall letter, the notification ask consignees to take the following actions: 1. If the user finds that PET reconstruction failing, the PET should be rescanned immediately to complete a new reconstruction. 2. If this occurs, please contact the UIH Solutions (UIHS) customer Contact Center at telephone numbers (855) 221-1552 (toll free) or (832) 699-2799 (direct) to report the issue.
Initiated
2019-08-26
Root cause
Software design
Terminated
2022-05-18
Distribution
US Nationwide distribution in the state TX.
Recall number
Z-1708-2020
Recall status
Terminated
CDRH record ID
180561
FDA record created
2020-04-15
Recall event number
85340

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗