Secondary collimator filters may be missing on certain x-ray units. It was discovered that the secondary collimator filter on some OEC 9600 C-arms was omitted when a collimator assembly was replaced. This would reduce the filtration and therefore the Half-Value-Layer (HVL) of the X-ray beam, and may result in non-compliance with respect to Maximum Entrance Exposure Rate (EER) unless the maximum EE
Source facts
Action
Domestic customer notification letters were sent via certified mail on 07/02/2008. They were advised to have the Maximum EER assessed and to discontinue use if the maximum limits were exceeded.
Initiated
2008-07-02
Root cause
Equipment maintenance
Terminated
2009-01-20
Distribution
Nationwide Distribution including states of CA, FL, MS, and TX.
Recall number
Z-1709-2008
Recall status
Terminated
CDRH record ID
73590
FDA classified
2008-09-20
Recall event number
49483