Z-1709-2019: Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The Alinity c Cuvettes are used on the Al

Recall record Z-1709-2019 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1709-2019
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
Jun 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.

Source facts

Action
On April 15, 2019, Abbott sent Product Correction Urgent - Immediate Action Required letters to all impacted Customers. US customers were sent these letters via FedEx priority overnight express while customers outside the US were notified via mailings or visits. The letters sent to US consignees were accompanied by Customer Reply form which were to be completed and sent back to Abbott.
Initiated
2019-04-15
Root cause
Under Investigation by firm
Terminated
2020-10-13
Distribution
US Nationwide Distribution- AL, AR,CA, FL, GA, IL, LA, MA, ME, MN, MO, NJ, NY, OK, SC, TN, TX, UT, & VA WORLDWIDE - Albania, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, Norway, Pakistan, Philippines, Poland, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Turkey, United Kingdom, and Vietnam.
Recall number
Z-1709-2019
Recall status
Terminated
CDRH record ID
172169
FDA record created
2019-06-02
Recall event number
82644

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗