Open original FDA record ↗Recall record
- FDA identifier
- Z-1710-2014
- Company in FDA record
- Leica Microsystems, Inc.
- FDA classification date
- Jun 4, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.
Source facts
- Action
- Leica sent an Urgent Field Safety Correction Notice letter dated January 10, 2014 was sent ot all affected customers. The notice included instructions for recipients to: 1) destroy any unused or partially used NCL-L-CD15 (lots 6015073 and 6017936) reagent or confirm that the products are no longer in stock; 2) indicate on the attached Field Safety Correction Notice Acknowledgement Fom which action was taken; and, 3) contact the local Leica represetnative immediately if a replacement is required. Customers with any concerns about this notice can contact Leica via e-mail at LMGRA@leica-microsystems.com.
- Initiated
- 2014-01-10
- Root cause
- Incorrect or no expiration date
- Terminated
- 2017-09-05
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, Romania, Russia, Saudi Arabia, Serbia, South Africa, Sweden, Taiwan, Tunisia, Turkey, and United Kingdom
- Recall number
- Z-1710-2014
- Recall status
- Terminated
- CDRH record ID
- 126925
- FDA classified
- 2014-06-04
- Recall event number
- 68090