Skip to content

Public FDA record

Z-1710-2020: Lunar iDXA Advance (iDXA with enCORE v16) bone densitometer

Recall record Z-1710-2020 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1710-2020
Company in FDA record
GE Healthcare, LLC
FDA classification date
Apr 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.

Source facts

Action
GE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from the measure screen of the enCORE software. GE Healthcare will correct all affected Lunar iDXA systems by inspecting and replacing, if necessary, the iDXA control panel at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2019-12-03
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-1710-2020
Recall status
Open, Classified
CDRH record ID
181018
FDA classified
2020-04-24
Recall event number
85466

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.