Open original FDA record ↗Recall record
- FDA identifier
- Z-1711-2017
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Mar 15, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.
Source facts
- Action
- An Urgent Medical Device Recall letter was sent on 3/7/17 to inform customers that Beckman Coulter has determined through internal investigation that a circuit board that controls motor functions in the instrument systems listed above can affect device positioning on the pipettor and incubator belt while the instrument is running. Customers are informed of the impact, action to be taken, and the resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
- Initiated
- 2017-03-07
- Root cause
- Software Design Change
- Terminated
- 2018-03-01
- Distribution
- Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe
- Recall number
- Z-1711-2017
- Recall status
- Terminated
- CDRH record ID
- 153847
- FDA classified
- 2017-03-15
- Recall event number
- 76688