Open original FDA record ↗Recall record
- FDA identifier
- Z-1713-2014
- Company in FDA record
- Arrow International Inc
- FDA classification date
- Jun 4, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Arrow is recalling the Arrow Percutaneous Sheath Introducer Kits due to dilator tip damage noted during internal inspection. Dilator tip damage has the potential to result in vessel damage.
Source facts
- Action
- Teleflex/Arrow International sent an Urgent Medical Device Recall Notification letter dated April 17, 2014, to distributors and end users. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any affected product and to complete the enclosed Recall Acknowledgement Form and fax it to 1-800-343-2935 to return product. Customers would be contacted by a customer service representative for a Return Goods Authorization (RGA) Numbers and would be provided instructions for the return of the product. Customers should complete the form whether they have affected product or not. Customers with questions were instructed to call 1-800-343-2935. For questions regarding this recall call 610-378-0131.
- Initiated
- 2014-04-14
- Root cause
- Process control
- Terminated
- 2016-10-14
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada.
- Recall number
- Z-1713-2014
- Recall status
- Terminated
- CDRH record ID
- 126904
- FDA classified
- 2014-06-04
- Recall event number
- 68073