Reason stated in the source
Expiration date discrepancy. Marked as 5 years. Expiration date should be 3 years.
Source facts
- Action
- On 2/28/2020, Cooper Surgical issued a "Medical Device Field Safety Notice" for one lot number (250887) of the Wallace Intra-Uterine Insemination Catheters due to a discrepancy in the expiration dates on the individual pouches, inner and outer boxes. The notice was sent to the Japan authority (PMDA) and affected customers in both English and Japanese. The recall notice states that the affected lot (1170 units) of product manufactured in August 2018 have not resulted in any adverse events and the product is acceptable for use until the expiration date on August 23, 2021. Customers that prefer to return the product or have no product on hand were provided instructions to complete the Acknowledgement and Receipt form.
- Initiated
- 2020-02-28
- Root cause
- Labeling Change Control
- Terminated
- 2020-08-11
- Distribution
- International distribution to Japan only.
- Recall number
- Z-1713-2020
- Recall status
- Terminated
- CDRH record ID
- 180163
- FDA record created
- 2020-04-15
- Recall event number
- 85139