Open original FDA record ↗Recall record
- FDA identifier
- Z-1714-2014
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Jun 4, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.
Source facts
- Action
- Toshiba America Medical Systems, Inc. initiated this recall by sending recall notifications via USPS return receipt mail. The recall notification letter, dated May 12, 2014, titled "URGENT: MEDICAL DEVICE CORRECTION", informed customers of the following: device information, serial number, SID, reason for reason, purpose of the notification, corrective action, request to customers, contact information, and customer reply form. Customers with questions regarding this letter were instructed to call (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968. For questions regarding this recall call 714-730-5000.
- Initiated
- 2014-05-12
- Root cause
- Other
- Terminated
- 2014-10-02
- Distribution
- Nationwide Distribution including CA and IN
- Recall number
- Z-1714-2014
- Recall status
- Terminated
- CDRH record ID
- 127470
- FDA classified
- 2014-06-04
- Recall event number
- 68363