Open original FDA record ↗Recall record
- FDA identifier
- Z-1714-2015
- Company in FDA record
- Advanced Sterilization Products
- FDA classification date
- Jun 5, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
Source facts
- Action
- A customer notification letter dated 5/7/15 will be sent to all customers on 5/13/15 to inform them that ASP is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels and subsequently pass the cycle without cancellation if there is a disconnected scope. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the attached Business Reply Card (double-sided) via mail or by fax to Stericycle: 888-879-8144. Customers with questions regarding the business reply card are instructed to contact Stericycle at 888-345-4671 and reference event #7781.
- Initiated
- 2015-05-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-07-06
- Distribution
- Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.
- Recall number
- Z-1714-2015
- Recall status
- Terminated
- CDRH record ID
- 136590
- FDA classified
- 2015-06-05
- Recall event number
- 71209