Z-1714-2019: Prometra Programmable Pump, Cat. No. 11827 For intrathecal infusion.

Recall record Z-1714-2019 for Flowonix Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1714-2019
Company in FDA record
Flowonix Medical Inc
FDA record creation date
Jun 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.

Source facts

Action
On March 21, 2019, the firm issued Urgent Recall Notices to customers to inform them of the product issue. In May 2019, the firm sent updated communications to customers: Healthcare providers who have Clinician Programmers with SW 2.00.29 were sent a follow-up recall letter. This letter included the information that the pump stop following "Error Code 115" will only occur after the Bridge or Demand Bolus is complete, which could take up to seventy-two (72 hours). In addition, in the event of any cause of an "Error Code 115", the pump sounds a one-beep alarm and then shuts down. This behavior differs from the Instructions for Use, which state that the pump will sound three long (half second) beeps every 30 minutes for all critical errors. As it is unlikely that the patient will notice the single beep, there may be a significant delay in the patient becoming aware that the pump has stopped delivering therapy which may in turn delay resumption of drug therapy. Healthcare providers who have Clinician Programmers with any other version of the software and the implant pumps received a Corrective Action Notice. This notice included the information from the preceding paragraph, noting the discrepant alarm behavior for "Error Code 115" could be missed by the patient. Customers with the new Prometra Clinician Programmer software (v. 2.00.29) were asked to do the following: 1. STOP using the Periodic Flow or Multiple Rates flow mode functions in combination with a Bridge or Demand Bolus using the Clinician Programmer with Software Version 2.00.29. The error only occurs when these functions are used together. a. In the event that a Bridge or Demand Bolus is needed, convert the patient to Constant Flow Mode, then program the Bridge or Demand Bolus as needed (instructions for use PL-11803-04, page 30). Once the bolus is complete, revisit the patient or have him/her return to your clinic for programming back into their original flow mode. 2. IDENTIFY
Initiated
2019-03-21
Root cause
Software change control
Terminated
2020-06-05
Distribution
US nationwide distribution.
Recall number
Z-1714-2019
Recall status
Terminated
CDRH record ID
172978
FDA record created
2019-06-03
Recall event number
82586

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗