Open original FDA record ↗Recall record
- FDA identifier
- Z-1715-2020
- Company in FDA record
- Smiths Medical ASD, Inc.
- FDA classification date
- Apr 15, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Portex Pro-Vent and LineDraw Arterial Blood Sampling syringes may have broken luers.
Source facts
- Action
- The firm initiated the recall by letter on January 07, 2020. The "Urgent Medical Device Recall Notice", identified the removal of devices and instructed the customer to quarantine the affected products. The Notice provided the accompanying Recall Response Form to be completed and returned within 10 days of receipt to smithmedical4136@stericycle.com. The Notice details that the affected product must be returned to Stericycle for processing. The Notice instructed the distributor to notify their customers and to provide the customer a copy of the Recall Notice and Response Form. The Notice provided the contact email smithmedical4136@stericycle.com and phone number of 8669188738 for assistance.
- Initiated
- 2020-01-07
- Root cause
- Process control
- Terminated
- 2020-12-14
- Distribution
- USA: AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI,MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA. Int'l: CA, DE, AU.
- Recall number
- Z-1715-2020
- Recall status
- Terminated
- CDRH record ID
- 179021
- FDA classified
- 2020-04-15
- Recall event number
- 84681