Skip to content

Public FDA record

Z-1716-2020: Portex ProVent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes

Recall record Z-1716-2020 for Smiths Medical ASD, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1716-2020
Company in FDA record
Smiths Medical ASD, Inc.
FDA classification date
Apr 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lots of Smiths Medical Portex Pro-Vent and LineDraw Arterial Blood Sampling syringes may have broken luers.

Source facts

Action
The firm initiated the recall by letter on January 07, 2020. The "Urgent Medical Device Recall Notice", identified the removal of devices and instructed the customer to quarantine the affected products. The Notice provided the accompanying Recall Response Form to be completed and returned within 10 days of receipt to smithmedical4136@stericycle.com. The Notice details that the affected product must be returned to Stericycle for processing. The Notice instructed the distributor to notify their customers and to provide the customer a copy of the Recall Notice and Response Form. The Notice provided the contact email smithmedical4136@stericycle.com and phone number of 8669188738 for assistance.
Initiated
2020-01-07
Root cause
Process control
Terminated
2020-12-14
Distribution
USA: AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI,MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA. Int'l: CA, DE, AU.
Recall number
Z-1716-2020
Recall status
Terminated
CDRH record ID
179023
FDA classified
2020-04-15
Recall event number
84681

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗