Z-1717-2020: Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or U

Recall record Z-1717-2020 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1717-2020
Company in FDA record
Sysmex America, Inc.
FDA record creation date
Apr 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.

Source facts

Action
The firm initiated the recall by letter on January 21, 2020. The "Product Notification", identified the Field Correction and instructed the customer to use the kit to direct fluid safety away from the conveyor electrical parts to address this potential hazard. During the next service visit, this kit will be installed by your Sysmex Field Service Engineer. The Notification also instructs the customer to: 1. Ensure that the surface of the sample, in the two, is more than away from the opening of the two. 2. Do not store samples on or above the analyzer. 3. if the sample transport conveyor unexpectedly stop, freezes, or jams, immediately remove all samples racks from the conveyor, before reinitializing or resetting analyzer or conveyor. 4. In the event of a sample spill on the conveyor, turn off the power to the equipment immediately and clean up any visible spills. If a spill occurs prior to the installation of the spill countermeasure kit, contact the Sysmex Technical Assistance Center (TAC). A Sysmex representative can explain how to properly clean the spill or will dispatch a Field Service Engineer. 5. Please distribute this Product Notification as appropriate to your laboratory staff. 6. File this Product Notification as a part of your laboratory's Quality System as required. the notice provided the contact information of the Technical Assistance Center at 1 888 879 7639 in the U.S., or at 1 888 679 7639 in Canada.
Initiated
2020-01-21
Root cause
Process control
Terminated
2023-01-30
Distribution
USA: FL, VT, IL.
Recall number
Z-1717-2020
Recall status
Terminated
CDRH record ID
179273
FDA record created
2020-04-15
Recall event number
84759

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗