Open original FDA record ↗Recall record
- FDA identifier
- Z-1718-2014
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Jun 4, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.
Source facts
- Action
- Intuitive sent an Urgent Medical Device Correction letter dated May 13, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. All personnel who use the device should be notified of the issue and the acknowledgement form completed and returned. Intuitive reps will visit each affect sites to install a replacement PSC and ensure that it is properly integrated with the Da Vinci. For questions contact Intuitive Surgical at 800-876-1310 Option 3 (Monday - Friday 6 am to 5 pm PST).
- Initiated
- 2014-05-09
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-06-04
- Distribution
- Medical facilities in TX and FL
- Recall number
- Z-1718-2014
- Recall status
- Terminated
- CDRH record ID
- 127460
- FDA classified
- 2014-06-04
- Recall event number
- 68358