Open original FDA record ↗Recall record
- FDA identifier
- Z-1719-2015
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Jun 5, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Memory can become corrupted when creating a fused study via drag and drop in Leksell GammaPlan 10.2.
Source facts
- Action
- The firm, Elekta, sent an "URGENT IMPORTANT FIELD SAFETY NOTICE 100-01-102-015" to all customers on 6/1/2015. The notice described the product, problem and actions to be taken. The customers were instructed to not use drag-and-drop to fuse images as a solution until a permanent fix can be developed and implemented on all affected devices. A solution is in development that will resolve the issue and is expected to be released September 2015. Elekta Service will then work with customers to implement on affected devices - estimated 6 month completion. Customers were also instructed to complete and return the IMPORTANT FIELD SAFETY NOTICE ACKNOWLEDGEMENT form to your local Elekta Office or Representative as soon as possible and within 30 days at the latest. If you have any queries about this Notice, please contact your local Elekta representative or the Global Post market Surveillance Manager at 770-300-9725 or email: Elekta_Global_PMS@elekta.com.
- Initiated
- 2015-06-01
- Root cause
- Software in the Use Environment
- Terminated
- 2016-01-12
- Distribution
- Worldwide Distribution: US (nationwide) including states: IL, OH, PA, TX, and WI; and countries of: Germany, Japan and United Kingdom.
- Recall number
- Z-1719-2015
- Recall status
- Terminated
- CDRH record ID
- 137250
- FDA classified
- 2015-06-05
- Recall event number
- 71295