Open original FDA record ↗Recall record
- FDA identifier
- Z-1724-2015
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Jun 8, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Recapper Caps may fall off of, or be knocked from sample tubes placed in the storage racks of entry/exit modules. The gripper may grab a tube at the wrong height and drop the tube in the Rack Entry/Exit or Rack Exit module. These situations may result in splashing of bio-hazardous fluid. Sample cross-contamination may occur with an uncapped tube in the rack.
Source facts
- Action
- On 5/06/2015, a customer letter (Ref. CL2015-092) with recommendation to prevent this issue was sent to all customers in the US with enGen Systems via FedEx or US Priority Mail (for Post Office Box addresses) or by email to subscribers of ORTHO PLUS e-Communications. Foreign affiliate consignees were notified by e-mail on 5/06/2015 informing them of the issue. For questions, please call Customer Technical Services at 1-800-421-3311 at any time.
- Initiated
- 2015-05-06
- Root cause
- Component change control
- Terminated
- 2018-05-25
- Distribution
- Worldwide distribution. US Nationwide, Australia, Bermuda, Brazil, Canada, Chile, China, France, India, Italy, Norway, Portugal, Spain, Sweden, and the United Kingdom.
- Recall number
- Z-1724-2015
- Recall status
- Terminated
- CDRH record ID
- 136349
- FDA classified
- 2015-06-08
- Recall event number
- 71174