Z-1725-2025: Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002A

Recall record Z-1725-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1725-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
May 9, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A cybersecurity vulnerability was discovered through internal testing.

Source facts

Action
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to healthcare providers and patients beginning on 04/07/2025 via USPS first class mail. The notice explained that Baxter was voluntarily initiating a permanent removal of all Life 2000 ventilators currently in use due to a cybersecurity issue and requested the following: Actions to be Taken by DME Providers: 1. Notify your patients using Life2000 ventilators of this recall. 2. Work with patients and prescribers to identify an alternate therapy option. 3. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. Actions to be Taken by Healthcare Providers: 1. Locate and discontinue use of all Life2000 ventilators and compressors within your facility. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors at your facility. Actions to be Taken by Distributors: 1. Please disseminate this information and conduct a user-level recall of the affected product that you distributed to customers. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. If you suspect that a device has been exposed to unauthorized personnel, contact Baxter Acute Care Customer Service team. The team can help verify whether the device settings are correct and if the device is performing as expected. The team can be reached by email at HRC_ACCS_Web@baxter.com, or by phone at 800-426-4224, option 2, then option 1. For questions regarding this communication, contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, between the hours of 8:00 am and 6:00 pm Central Time, Monday through Friday.
Initiated
2025-04-07
Root cause
Software in the Use Environment
Distribution
Worldwide distribution.
Recall number
Z-1725-2025
Recall status
Open, Classified
CDRH record ID
213249
FDA classified
2025-05-09
Recall event number
96611

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗