Skip to content

Public FDA record

Z-1727-2020: Nikon Metrology XTH cabinet x-ray Systems which includes variations STH225, XTH225ST, and MCT225.

Recall record Z-1727-2020 for Nikon Metrology. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1727-2020
Company in FDA record
Nikon Metrology
FDA record creation date
May 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The manufacturer identified a risk that their products might fail to comply with the cabinet x-ray system performance standard limit on x-ray emissions [21 CFR 1020.40(c)(1)] due to an insufficient safety margin for the means of attachment of a piece of interior x-ray shielding. Double sided industrial tape was used for attachment - there have been only 2 instances of detachment and although the piece of shielding detached it remained mostly in place supported by wires that pass through that piece of shielding. The repair is accomplished by adding a bracket to fasten the piece of shielding securely to the inside of the x-ray system. Mfr reports that no systems were found to have exceeded the emission limit when checked prior to repair.

Source facts

Action
Nikon sent Notification letters to affected customers. The letter identified the Description of the Possible Defect, Potential Health Risk from the Defect, and actions to be taken. For question contact the service department via email at xray-support.nm-us@nikon or phone at (810)220-4360.
Initiated
2019-12-23
Root cause
Radiation Control for Health and Safety Act
Terminated
2024-06-13
Distribution
US Nationwide Distribution
Recall number
Z-1727-2020
Recall status
Terminated
CDRH record ID
181102
FDA record created
2020-05-22
Recall event number
85498

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.