Z-1728-2011: Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-001

Recall record Z-1728-2011 for QIAGEN Gaithersburg, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1728-2011
Company in FDA record
QIAGEN Gaithersburg, Inc.
FDA record creation date
Jun 26, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The software component of a diagnostic medical device may be defective and could cause incorrect patient values.

Source facts

Action
The firm, Qiagen, sent a "Notice of Product Field Action" letter dated January 12, 2011 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue the process of running multiple assay protocols on a single microplate to eliminate the potential for inaccurate patient result reporting, and to complete and return the acknowledgment section via fax to QIAGEN 661-702-3854 or scan the document and email to techservice-na@qiagen.com. The firm informed users that they will provide further detailed instructions for a workaround and software fix. A QIAGEN representative will contact the customers who indicated that they run multiple assays on a plate and are currently using the HC2 System Software Version 3.0 (Suite 4.0), to assist in the identification of those patients that may have been affected by the software defect. If you have any questions or concerns, please contact Qiagen Technical Services at 800-344-3631, Option 2.
Initiated
2011-01-12
Root cause
Software design
Terminated
2011-11-14
Distribution
The recalled product was sold to medical laboratories nationwide and in Canada.
Recall number
Z-1728-2011
Recall status
Terminated
CDRH record ID
97490
FDA record created
2015-06-26
Recall event number
57814

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗