Unexpected Movement: if the stereotactic motion disable function is turned on, the couch can be moved via float mode unexpectedly.
Source facts
Action
Varian Medical Systems issued an "Urgent Medical Device Correction" dated May 29, 2009 distributed to all affected users, as well as Varian Sales, Marketing and Service organizations, informing them of the issue. Consignees were informed that a Varian Service representative will be in contact to schedule the proper correction.
For further questions, contact your local Varian Customer Support District Manager.
Initiated
2009-05-29
Root cause
Device Design
Terminated
2011-05-18
Distribution
Worldwide Distribution.
Recall number
Z-1729-2009
Recall status
Terminated
CDRH record ID
83530
FDA classified
2009-08-27
Recall event number
52636