Z-1729-2014: HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with

Recall record Z-1729-2014 for Cenorin, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1729-2014
Company in FDA record
Cenorin, LLC
FDA classification date
Jun 6, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.

Source facts

Action
Cenorin sent a letter on April 24, 2014 informing their customers of the potential failure of the pump. This letter also explains the customer's option of upgrading their entire HLD System to the new cleaning fluid delivery system or to continue purchasing a six month PM kit that has a replacement pump added. Customers are instructed to complete and return an enclosed response form within 15 days and to email Jamie Wright at jwright@cenorin.com or call 800-426-1042, extension 224 for questions.
Initiated
2014-04-24
Root cause
Equipment maintenance
Terminated
2015-06-30
Distribution
Distribution USA Nationwide
Recall number
Z-1729-2014
Recall status
Terminated
CDRH record ID
127201
FDA classified
2014-06-06
Recall event number
68237

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗