Open original FDA record ↗Recall record
- FDA identifier
- Z-1729-2015
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Jun 8, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.
Source facts
- Action
- Ortho Clinical Diagnostics distributed correction notification letters dated April 6, 2015, and response forms to their customers via courier service. All urine patient results that require pretreatment was advised to refer to the notice of correction for instructions. Correction notices should also be posted beside the device user documentation for easy reference. Customers were advised to complete and return the Confirmation of Receipt before April 20, 2015. Customers with questions may contact the Customer Technical Service at 1- 800-421-3311 anytime.
- Initiated
- 2015-04-06
- Root cause
- Software design
- Terminated
- 2018-05-25
- Distribution
- Worldwide Distribution - US Nationwide, Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.
- Recall number
- Z-1729-2015
- Recall status
- Terminated
- CDRH record ID
- 137113
- FDA classified
- 2015-06-08
- Recall event number
- 70918