Z-1732-2026: eTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150

Recall record Z-1732-2026 for Civco Medical Instruments Co. Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1732-2026
Company in FDA record
Civco Medical Instruments Co. Inc.
FDA classification date
Apr 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 3/2/26 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any eTRAX Needle Sensor for Aurora Trackers and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory, or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any eTRAX Needle Sensor for Aurora Trackers remaining in your inventory to assist in our reconciliation process. Response forms can also be sent via email to distorder@civco.com. For questions, contact CIVCO at # 800-445-6741, or 319-248-6757. Type of Action by the Company: This notice is being communicated to all customers who have purchased an eTRAX Needle Sensor for Aurora Trackers from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level.
Initiated
2026-03-02
Root cause
Software change control
Distribution
US States: OH, PA, WA. China.
Recall number
Z-1732-2026
Recall status
Open, Classified
CDRH record ID
218957
FDA classified
2026-04-01
Recall event number
98513

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗