Z-1734-2019: GETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 7

Recall record Z-1734-2019 for Maquet Cardiovascular Us Sales, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1734-2019
Company in FDA record
Maquet Cardiovascular Us Sales, Llc
FDA classification date
Jun 4, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

User may suffer an electric shock when coming in contact with the exposed electronic components

Source facts

Action
Maquet /Getinge issued URGENT MEDICAL DEVICE CORRECTION Notice on 5/6/19, identified issued, health risk and action to take: Per the IFU, section 8.1, a visual and functional check is required to be performed by a trained person before operation; If the plastic socket of the cold-device plug is pulled out do not try to put it back into the housing and if there are any exposed electronic components: do not touch them. Call your in-house technical service and arrange the repair of the table. Per IFU section 8.5 a defective product may not be used and may not be repaired by yourself. Complete/sign and return the MEDICAL DEVICE RECALL CORRECTION RESPONSE FORM. Questions contact Technical Support Department at 1-888-627-8383 (press option 3 then option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. Expanded Recall: Maquet/Getinge issued letters on 9/24/20 to consignees via FedEx 2 day Delivery with Signature Proof of Delivery (SPOD).
Initiated
2019-05-06
Root cause
Process control
Terminated
2022-07-06
Distribution
Worldwide - US Nationwide Distribution Foreign: ALGERIA ANGOLA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BOLIVIA BOTSWANA BOSNIA AND HERZEGOVINA BRAZIL BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK ECUADOR EGYPT FINLAND FRANCE GABON GERMANY GREECE HONG KONG HONDURAS HUNGARY ICELAND INDIA IRAN IRELAND ISRAEL ITALY JAPAN JORDAN KUWAIT LATVIA LEBANON LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVA MONTENEGRO MOROCCO MYANMAR NETHERLANDS NAMIBIA NORWAY NEW ZEALAND PAKISTAN PARAGUAY PERU PHILIPPINES POLAND PORTUGAL ROMANIA SAUDI ARABIA SERBIA SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TURKEY TURKMENISTAN UKRAINE UNITED ARAB EMIRATES URUGUAY UNITED KINGDOM VIETNAM
Recall number
Z-1734-2019
Recall status
Terminated
CDRH record ID
172719
FDA classified
2019-06-04
Recall event number
82792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗