Open original FDA record ↗Recall record
- FDA identifier
- Z-1734-2019
- Company in FDA record
- Maquet Cardiovascular Us Sales, Llc
- FDA classification date
- Jun 4, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
User may suffer an electric shock when coming in contact with the exposed electronic components
Source facts
- Action
- Maquet /Getinge issued URGENT MEDICAL DEVICE CORRECTION Notice on 5/6/19, identified issued, health risk and action to take: Per the IFU, section 8.1, a visual and functional check is required to be performed by a trained person before operation; If the plastic socket of the cold-device plug is pulled out do not try to put it back into the housing and if there are any exposed electronic components: do not touch them. Call your in-house technical service and arrange the repair of the table. Per IFU section 8.5 a defective product may not be used and may not be repaired by yourself. Complete/sign and return the MEDICAL DEVICE RECALL CORRECTION RESPONSE FORM. Questions contact Technical Support Department at 1-888-627-8383 (press option 3 then option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. Expanded Recall: Maquet/Getinge issued letters on 9/24/20 to consignees via FedEx 2 day Delivery with Signature Proof of Delivery (SPOD).
- Initiated
- 2019-05-06
- Root cause
- Process control
- Terminated
- 2022-07-06
- Distribution
- Worldwide - US Nationwide Distribution Foreign: ALGERIA ANGOLA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BOLIVIA BOTSWANA BOSNIA AND HERZEGOVINA BRAZIL BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK ECUADOR EGYPT FINLAND FRANCE GABON GERMANY GREECE HONG KONG HONDURAS HUNGARY ICELAND INDIA IRAN IRELAND ISRAEL ITALY JAPAN JORDAN KUWAIT LATVIA LEBANON LITHUANIA MACEDONIA MALAYSIA MAURITIUS MEXICO MOLDAVA MONTENEGRO MOROCCO MYANMAR NETHERLANDS NAMIBIA NORWAY NEW ZEALAND PAKISTAN PARAGUAY PERU PHILIPPINES POLAND PORTUGAL ROMANIA SAUDI ARABIA SERBIA SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND SYRIA TAIWAN TANZANIA THAILAND TURKEY TURKMENISTAN UKRAINE UNITED ARAB EMIRATES URUGUAY UNITED KINGDOM VIETNAM
- Recall number
- Z-1734-2019
- Recall status
- Terminated
- CDRH record ID
- 172719
- FDA classified
- 2019-06-04
- Recall event number
- 82792