Open original FDA record ↗Recall record
- FDA identifier
- Z-1735-2019
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Jun 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Etching on the tulip head incorrectly states "24MM" instead of the correct "28MM" marking.
Source facts
- Action
- On November 14, 2018, the firm sent an "Urgent Medical Device Recall" letters dated November 14, 2018, to its affected customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete the included Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-447-3625 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
- Initiated
- 2018-11-14
- Root cause
- Process control
- Terminated
- 2020-05-12
- Distribution
- US Distribution to CA, NC, and NY; and International to Germany.
- Recall number
- Z-1735-2019
- Recall status
- Terminated
- CDRH record ID
- 172921
- FDA record created
- 2019-06-05
- Recall event number
- 82856