Z-1736-2011: Access Total BhCG (BhCG) Assay Part Number 33500 The Access Total BhCG assay is a paramagnetic particle,

Recall record Z-1736-2011 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1736-2011
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Mar 21, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Beckman Coulter has confirmed that non-reproducible falsely elevated test results may occur when using the Access Total BhCG assay. These elevated results are often attributed to pre-analytical factors and are particularly noted at the low end of the analytical measuring range. The role of preanalytical factors in laboratory testing has been described in a variety of published literature.

Source facts

Action
Beckman Coulter sent an Urgent Product Corrective Action letter dated November 11, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to:. " Evaluate current laboratory protocols to ensure the necessary procedures are in place for identifying and mitigating falsely elevated test results. " Follow the sample handling instructions provided in the Access Total hCG Instruction for Use (IFU). " Review the enclosed pre-analytical sample handling information bulletins: 1. The Role of Preanalytical Factors in Chemistry and Immunoassay Testing (P/N A34880A). 2. Sample Handling Makes a Difference (P/N A28226A). " At the discretion of the laboratory director, any total hCG result that is questioned due to lack of clinical correlation should be repeated. NOTE: Due to the clinical utility of the Access Total hCG assay as a biomarker of pregnancy, Beckman Coulter feels it is unnecessary to require a review of all previously obtained Access Total hCG test results. The Access Total hCG assay is intended as a marker for pregnancy at the time of testing. A discrepant or questionable result would have been brought to the attention of the laboratory. The pregnancy status of a patient at this point, post testing, would be known to the patient and the physician. " Please share this information with your laboratory staff and your phlebotomy director and/or phlebotomy staff and retain this notification as part of your laboratory Quality System documentation. " Complete and return the enclosed response form within ten days so they can be assured that this important notification has been received. If you have any questions regarding this notification, please contact the Customer Support Center on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support or call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada
Initiated
2010-10-20
Root cause
Pending
Terminated
2014-09-02
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Algeria, Angola, Antigua and Barbuda, Argentina, Armenia, Australia, Bangladesh, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Congo, Croatia, Czech Republic, Djibouti, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Philippines, Poland, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Zambia, Zimbabwe.
Recall number
Z-1736-2011
Recall status
Terminated
CDRH record ID
96242
FDA classified
2011-03-21
Recall event number
57443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗