One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.
Source facts
Action
SpineFrontier issued a Recall Notification to the affected distributor via telephone. The distributor was instructed to return the affected product for a replacement product.
Initiated
2011-12-30
Root cause
Vendor change control
Terminated
2014-01-08
Distribution
USA Nationwide Distribution including the state of TX
Recall number
Z-1737-2013
Recall status
Terminated
CDRH record ID
117587
FDA classified
2013-07-16
Recall event number
64989