Skip to content

Public FDA record

Z-1737-2022: Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per pack

Recall record Z-1737-2022 for Technomed Europe. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1737-2022
Company in FDA record
Technomed Europe
FDA record creation date
Sep 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.

Source facts

Action
The recalling firm issued letters dated 8/9/2022 that were flagged "MRI Safety Information." The letter provided an overview of the issue, the affected devices and lot numbers, a description of the issue including the MRI Safety Information that was added to the Instructions for Use, the hazard, and the actions to be taken. The actions included the consignee is to read the Field Safety Notice, review the list of affected products, inform customers who received or will receive products from the affected lot numbers about the issue by forwarding a copy of the notice and the updated Instructions for Use to them, and complete and sign the enclosed acknowledgment form to indicate they have read the notice, understand its content, and will follow the instructions as intended. The form is to be returned via email.
Initiated
2022-08-09
Root cause
Labeling Change Control
Terminated
2024-04-11
Distribution
US Nationwide distribution in the states of CA, FL, NH, OH, and WA.
Recall number
Z-1737-2022
Recall status
Terminated
CDRH record ID
195598
FDA record created
2022-09-12
Recall event number
90733

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗