Z-1738-2008: Zimmer Spine Cyclone anterior cervical plate, 3 Level, 60MM (Ti-6Al-4V); REF 600-03-160.

Recall record Z-1738-2008 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1738-2008
Company in FDA record
Zimmer Inc.
FDA classification date
Sep 1, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The screw locking cap may fracture when the surgeon rotates it into the locked position.

Source facts

Action
Consignees were notified via voicemail on 3/26/08 and a letter was sent to all distributors on 3/27/08 informing them that the plate should no longer be used and that all lots are being recalled. A letter dated 6/18/08 was issued to implant physicians recommending patient monitoring. Contact Zimmer Spine at 1-952-857-5636 for assistance.
Initiated
2008-03-26
Root cause
Device Design
Terminated
2009-11-12
Distribution
Arizona, California, Florida, Georgia, Louisiana, Michigan, Minnesota, Montana, New Hampshire, North Carolina, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, Washington and Wisconsin.
Recall number
Z-1738-2008
Recall status
Terminated
CDRH record ID
69700
FDA classified
2008-09-01
Recall event number
47527

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗