There are 1-2 places on some of the multi-purpose kit blisters/trays in which there are holes and/or the material is so thin that holes are easily created. This would potentially cause the sterility of the products to be compromised.
Source facts
Action
Argon Medical Devices issued an "Urgent Medical Device Recall" notification via certified mail dated March 29, 2010 to distributors and end users. A Response Form was attached. Consignees were informed of the affected product and asked to quarantine and return all unused product to the firm.
For further information, contact Argon Medical Devices at 1-903-677-9375.
Initiated
2010-03-29
Root cause
Package design/selection
Terminated
2010-07-29
Distribution
Distributed to United States (IL, AL, NY, TX, OK, MI, NC, CA, OH, VA, KY, CO, LA, WA, MO, AND IA) and Canada.
Recall number
Z-1738-2010
Recall status
Terminated
CDRH record ID
90510
FDA classified
2010-06-02
Recall event number
55234