Haemonetics has received a report of smoke from the unit with a power supply failure.
Source facts
Action
Haemoscope sent letters to all of their U.S. accounts on January 24, 2011, via certified mail. The letters listed guidelines regarding proper use of the TEG 5000 System Power Supply to eliminate any potential risks such as electrical shorting or power supply damage. The information, which was printed in the letter, will also be available in an updated TEG User Manual soon to be released. A Caution label with installation instructions was included with the letter, to be applied to the back of the TEG unit. The accounts were requested to complete the enclosed acknowledgement form and fax it back to Haemoscope at 847-588-0455. Any questions were directed to 1-800-438-2834.
Initiated
2011-01-28
Root cause
Other
Terminated
2012-03-12
Distribution
Nationwide Distribution including CA, FL, MD, PA, OH, SC, TN, TX, and VA
Recall number
Z-1738-2011
Recall status
Terminated
CDRH record ID
97481
FDA classified
2011-03-21
Recall event number
57809