Open original FDA record ↗Recall record
- FDA identifier
- Z-1739-2013
- Company in FDA record
- St. Jude Medical
- FDA classification date
- Jul 16, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica
Source facts
- Action
- ST. Jude Medical sent an Important Medical Device Safety Information dated July 26, 2012 to affected customers. The letter identifed the affected product, problem and actions to be taken. For questions or concerns contact your ST. Jude Medical representative or St. Jude Medical Neuromodulation Division at 1-866-240-6741.
- Initiated
- 2012-07-26
- Root cause
- Device Design
- Terminated
- 2015-11-18
- Distribution
- Worldwide Distribution - USA Nationwide Including Puerto Rico and countries of: Argentina, Australia, Belgium, Brazil, Canada, China, Chile, Colombia, Ireland, Israel, India, Malaysia, Mexico, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland and United Kingdom.
- Recall number
- Z-1739-2013
- Recall status
- Terminated
- CDRH record ID
- 119936
- FDA classified
- 2013-07-16
- Recall event number
- 63109