Class II

Reason stated in the source

Acidified Concentrate Distribution System potential electrical safety hazard may result in electric shock, arching

Source facts

Action
Isopure initiated contact via telephone on 7/26/22 followed by Emails to consignees. .Letter states reason for recall, health risk and action to take: 1. Refer to Distribution List of ACDS units installed. 2. Perform an inspection of the ACDS device, associated wiring, cabling and cord. Provide a photograph if damage is found during inspection. 3. Video links are included in the inspection packet of the Isopure ACDS Instructions for Electrical Review and Preventive Maintenance of the ACDS and proper use of a torque screwdriver as a guide to inspecting electrical connections WI-50602. Please follow these guidance documents. Adhere to all safety information provided. 4. Fill out the Acknowledgement and Inspection Form upon completion of the inspection. Return the Acknowledgment form back to Isopure to Quality@isopure.com by September 30, 2022. 5. Always replace any depleted or blown fuses with the correct type and correct rating. 6. Follow the recommended preventive maintenance schedule for the device (refer to LA704262 ACDS Acidified Concentrate Distribution and Storage System User Manual). 7. Contact Isopure Customer/Technical Support at 502-967-7873 with any questions or concerns. Refer to: How -To-Video- Instructions ACDS Preventive Maintenance Instructions https://vimeo.com/732224767
Initiated
2022-07-26
Root cause
Device Design
Terminated
2024-05-31
Distribution
US Nationwide distribution.
Recall number
Z-1739-2022
Recall status
Terminated
CDRH record ID
195611
FDA classified
2022-09-13
Recall event number
90739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗