Potential for exoskeleton to lose lateral balance and cause patient to fall.
Source facts
Action
An URGENT SAFETY NOTICE dated 4/3/25 was mailed to consignees. The letter notifies consignees that WANDERCRAFT will be reinforcing their training program for the Atalante X and providing additional instructions by 5/1/25. Until the additional instructions can be provided, users are being instructed to only have certified Atalante X operators maintain the stability of the exoskeleton at all times by keeping at least one hand on the handles throughout the session. Consignees are to share the recall notification with users at their facility and retained for awareness.
Initiated
2025-04-03
Root cause
Use error
Distribution
Domestic: PA & NY.
Recall number
Z-1739-2025
Recall status
Open, Classified
CDRH record ID
213308
FDA classified
2025-05-08
Recall event number
96635